SCALE CALIBRATION (MODELS WITH A SCALE ONLY) 4 - 5
Step 4 - Calibration at LOAD
The Scale Display blinks either CAL25 or CAL11, depending on the units of measure programmed
(CAL25 = pounds; CAL11 = kilograms).
The diagnostic 7 -segment display scrolls C25Lb or C11g.
To continue with calibration:
1. Depending on the units of measure programmed, place 25 lb or 11 kg on the scale. Ideally these weights
should be grouped in the center of the scale (Figure 4-1).
Figure 4-1
(A) Center of Scale Location, and (B) Center of Scale Location with Platter Installed
IMPORTANT This step of the calibration procedure must be completed within 30 seconds, or 90 seconds for
Firmware Version before 3.4. (See Identifying Firmware Version.) If this time expires before
completing the step then the calibration procedure is exited.
A
B
Center of
Scale